SYD-261 Pensky Martens Closed Cup Flash Point tester Flash point
- メーカー
- Baoshishan
- JAN
- 7458621620566
- メーカー品番
- Baoshishan1167
- この商品のレビュー
- Amazonでのレビュー
- シェア
商品説明
BAOSHISHAN
Specifications
1.Power supply:AC (220?0%)V, 50Hz.
2. Heating device:The heating power is continuously adjustable from 0 W to 600W.
3. Heating rate:(1~12)degree/min;Controllable and adjustable.
4. Stirring rate:(1) Procedure A: (90~120)RPM(2) Procedure B: (250?0)RPM
5. Standard oil cup;(1) Inner diameter: 50.7mm~50.8mm.(2) Depth: 55.7mm~56.0mm(3) The scribed line depth of capacity of testing oil: 33.9mm~34.3mm.(4) Capacity of testing oil:about 70ml
6. Igniting device:(1) Igniting source: gas( or other civilian fuels, the same below )(2) Electric ignition. Gas flame diameter:3.2mm~4.8mm
7. Thermometers:Mercury-in-glass thermometer.
Specifications are as below:
(1)Scale -5degree~110degree, division 0.5degree.
(2)Scale 20degree~150degree, division 1 degree.
(3)Scale 90degree~370degree,division 2degree.
8. Ambient temperature:? 35degree
9. Relative humidity:? 85%
10.Maximum power consumption:650W
Note
All products have been checked and packaged in good condition before shipping.
After your order is confirmed,the product will be sent to your hands for 7-13 workdays.
If you have any question, please feel free to contact with us . We will try our best to give you a satisfactory solution.
特記事項
- 1.It is used to make determination of the closed cup flash point of the petroleum products.
- 2.SYD-261 Pensky Martens Closed Cup Flash Point tester Flash point This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester.
- 3.This instrument is designed as per GB/T261 and ASTM D93.
- 4.The heating power is continuously and steplessly adjustable. The power is shown visually by voltmeter
- 5.It has been chosen as the instrument to test the closed cup flash point by many metering and detecting institutes.